Modular versus nonmodular tibial inserts in total knee arthroplasty: what are the differences?
Although both modular and nonmodular tibial components in primary total knee arthroplasty (TKA) have shown excellent long-term survivorship (1-7), modularity in TKA has been a topic of ongoing debate over the past two decades. The authors of the article entitled “Twenty-Five-Years and Greater, Results After Nonmodular Cemented Total Knee Arthroplasty” should be congratulated for their contribution to the literature (1). The authors have reported the long-term results of nonmodular metal backed cemented tibial insert in 5,649 TKA procedures. The overall prosthesis survival rate was 94.2% at 25 years and 92.4% at 30 years. To the best of our knowledge, these results represent the longest reported follow up for nonmodular metal backed cemented TKAs.
The issue of whether the tibial component in TKA should be modular or not has continued since the inception of TKA. The advocates of modularity claim that the use of modular tibial TKA components has several advantages. It enables surgeons to select the tibial polyethylene thickness and gives them the opportunity to make intraoperative modifications after final tibial tray placement. Additional advantages in primary TKAs are the ease of implant placement and, in rare occasions, providing a better access to the back of the knee when cement extrusion is encountered. In revision cases, removing the tibial polyethylene at the beginning of the operative procedure helps improve soft tissue exposure and allow for better visibility and safe tibial and femoral component removal if indicated. In addition to that, modular tibial TKA provides an opportunity to the surgeon to perform simple isolated liner exchange rather than complete tibial component revision when treating osteolysis. Griffin et al. (8) have shown 84% success rate when isolated polyethylene exchange was performed for 68 patients who had wear and osteolysis after TKA. They also noted lack of progression of osteolytic lesions in the majority of cases at mean follow up period of 44 months. Callaghan et al. (9) have studied the utility of isolated liner exchange and bone grafting in cases with severe osteolysis around the knee in otherwise well fixed, aligned and stable TKAs. The authors found that liner exchange and bone grafting provided durable midterm results with extensive graft incorporation. The availability of this treatment option in modular TKA would preclude the need of removing a well-fixed tibial component and prevents potential substantial bone loss, reduces blood loss and leads to quicker rehabilitation and recovery. However, some authors have found that the clinical indication of isolated liner exchange is very rare and considered a theoretical advantage, while others found limited effectiveness of this treatment approach and reported high revision rates (25%) when isolated liner exchange was performed (10-12).
The drawback of modular tibial components in TKA is the fretting wear that occur between the interface of metal tibial tray and polyethylene (backside wear). This was proven to be a substantial problem in modular TKAs (13), particularly, when cementless tibial components with screw holes were implanted (14-16). This prompts surgeons to use cemented tibial components without screw holes aiming to reduce the incidence of backside wear and osteolysis. Nevertheless, these components may still generate wear and osteolysis (17,18). Retrieval studies have improved our knowledge regarding implant related factors on backside wear (19-21). It has been shown that using partial constrained peripheral polyethylene locking mechanism has better performance in preventing micromotion when compared to central polyethylene locking mechanisms(19). Furthermore, polished tibial trays were found to reduce the incidence of backside wear. Patient related factors also play a role on the rate of wear and osteolysis. Odland et al. (22) found that implanting modular TKAs in younger active population may increase the risk of wear and osteolysis. Backside wear in modular tibial TKAs remains a challenging problem despite the marketing efforts by the manufactures. Fortunately, wear-induced osteolysis in TKA is not a major cause of failure at long-term follow-up (23). One proposed strategy to decrease wear rates in TKA is using high cross-linked polyethylene liners. However, the available results are conflicting and the future long term follow up reports are required to provide insights on safety, potential complications and backside wear characteristic of these implants (23-25). Finally, some authors suggested that the drawbacks of modular tibial TKAs is not limited to backside wear but also creating another interface (between polyethylene and metal tray) that may potentially serve as a nidus for infection (26).
In addition to the excellent survivorship of nonmodular cemented metal backed and all polyethylene tibial components (1,5,7), these implants eliminate the possibility of motion at the polyethylene and tibial tray interface and therefore pose no risk of backside wear. Additional advantage of nonmodular TKAs is that it may decrease the inventory carried by hospitals when compared with that necessary for modular TKA implants (27). However, this may differ from specific implant system to another. The disadvantages of nonmodular TKA designs are the lack of intraoperative versatility offered by the modular TKA as well as the potential need of performing a complete tibial revision, in revision cases, since simple liner exchange is not an option. These revisions can be more destructive and cost-effective option when compared to isolated liner exchange only.
Taken together, both modular and nonmodular tibial components have advantages and drawbacks. Although some retrospective reports have favored the use nonmodular tibial TKA due to better survivorship results, prospective level I studies have shown similar functional and clinical outcomes when nonmodular all polyethylene cemented tibial design was compared to modular metal backed tibiae (7,26,28-31). The most important difference was found to be substantial reduction in cost (approximately $1,650 per case) when nonmodular all polyethylene cemented tibial implants were used (7). Currently, it appears that surgeon’s prior training and exposure and the available resources are the primary drives to choose between modular versus nonmodular TKAs. With the emerging transition to the “value-based care” and the rise of cost-effectiveness concerns, we anticipate that nonmodular cemented all polyethylene tibiae in TKAs will be broadly used in the United States in the upcoming years.
Acknowledgements
None.
Footnote
Conflicts of Interest: The authors have no conflicts of interest to declare.
References
- Ritter MA, Keating EM, Sueyoshi T, et al. Twenty-Five-Years and Greater, Results After Nonmodular Cemented Total Knee Arthroplasty. J Arthroplasty 2016;31:2199-202. [Crossref] [PubMed]
- Dixon MC, Brown RR, Parsch D, et al. Modular fixed-bearing total knee arthroplasty with retention of the posterior cruciate ligament. A study of patients followed for a minimum of fifteen years. J Bone Joint Surg Am 2005;87:598-603. [Crossref] [PubMed]
- Parsch D, Kruger M, Moser MT, et al. Follow-up of 11-16 years after modular fixed-bearing TKA. Int Orthop 2009;33:431-5. [Crossref] [PubMed]
- Vessely MB, Whaley AL, Harmsen WS, et al. The Chitranjan Ranawat Award: Long-term survivorship and failure modes of 1000 cemented condylar total knee arthroplasties. Clin Orthop Relat Res 2006.28-34. [Crossref] [PubMed]
- Ritter MA. The Anatomical Graduated Component total knee replacement: a long-term evaluation with 20-year survival analysis. J Bone Joint Surg Br 2009;91:745-9. [Crossref] [PubMed]
- Himanen AK, Belt EA, Lehto MU, et al. A comparison of survival of moulded monoblock and modular tibial components of 751 AGC total knee replacements in the treatment of rheumatoid arthritis. J Bone Joint Surg Br 2007;89:609-14. [Crossref] [PubMed]
- Voss B, El-Othmani MM, Schnur AK, et al. A Meta-Analysis Comparing All-Polyethylene Tibial Component to Metal-Backed Tibial Component in Total Knee Arthroplasty: Assessing Survivorship and Functional Outcomes. J Arthroplasty 2016;31:2628-36. [Crossref] [PubMed]
- Griffin WL, Scott RD, Dalury DF, et al. Modular insert exchange in knee arthroplasty for treatment of wear and osteolysis. Clin Orthop Relat Res 2007.132-7. [PubMed]
- Callaghan JJ, Reynolds ER, Ting NT, et al. Liner exchange and bone grafting: rare option to treat wear & lysis of stable TKAs. Clin Orthop Relat Res 2011;469:154-9. [Crossref] [PubMed]
- Babis GC, Trousdale RT, Morrey BF. The effectiveness of isolated tibial insert exchange in revision total knee arthroplasty. J Bone Joint Surg Am 2002;84-A:64-8. [Crossref] [PubMed]
- Bert JM, Reuben J, Kelly F, et al. The incidence of modular tibial polyethylene insert exchange in total knee arthroplasty when polyethylene failure occurs. J Arthroplasty 1998;13:609-14. [Crossref] [PubMed]
- Blumenfeld TJ, Scott RD. The role of the cemented all-polyethylene tibial component in total knee replacement: a 30-year patient follow-up and review of the literature. Knee 2010;17:412-6. [Crossref] [PubMed]
- Engh GA, Dwyer KA, Hanes CK. Polyethylene wear of metal-backed tibial components in total and unicompartmental knee prostheses. J Bone Joint Surg Br 1992;74:9-17. [PubMed]
- Peters PC Jr, Engh GA, Dwyer KA, et al. Osteolysis after total knee arthroplasty without cement. J Bone Joint Surg Am 1992;74:864-76. [Crossref] [PubMed]
- Lewis PL, Rorabeck CH, Bourne RB. Screw osteolysis after cementless total knee replacement. Clin Orthop Relat Res 1995.173-7. [PubMed]
- Berger RA, Lyon JH, Jacobs JJ, et al. Problems with cementless total knee arthroplasty at 11 years followup. Clin Orthop Relat Res 2001.196-207. [Crossref] [PubMed]
- Ries MD, Guiney W Jr, Lynch F. Osteolysis associated with cemented total knee arthroplasty. A case report. J Arthroplasty 1994;9:555-8. [Crossref] [PubMed]
- Griffin FM, Scuderi GR, Gillis AM, et al. Osteolysis associated with cemented total knee arthroplasty. J Arthroplasty 1998;13:592-8. [Crossref] [PubMed]
- Brandt JM, MacDonald SJ, Bourne RB, et al. Retrieval analysis of modular total knee replacements: factors influencing backside surface damage. Knee 2012;19:306-15. [Crossref] [PubMed]
- Berry DJ, Currier JH, Mayor MB, et al. Knee wear measured in retrievals: a polished tray reduces insert wear. Clin Orthop Relat Res 2012;470:1860-8. [Crossref] [PubMed]
- Conditt MA, Stein JA, Noble PC. Factors affecting the severity of backside wear of modular tibial inserts. J Bone Joint Surg Am 2004;86-A:305-11. [Crossref] [PubMed]
- Odland AN, Callaghan JJ, Liu SS, et al. Wear and lysis is the problem in modular TKA in the young OA patient at 10 years. Clin Orthop Relat Res 2011;469:41-7. [Crossref] [PubMed]
- Sakellariou VI, Sculco P, Poultsides L, et al. Highly cross-linked polyethylene may not have an advantage in total knee arthroplasty. HSS J 2013;9:264-9. [Crossref] [PubMed]
- Lachiewicz PF, Geyer MR. The use of highly cross-linked polyethylene in total knee arthroplasty. J Am Acad Orthop Surg 2011;19:143-51. [Crossref] [PubMed]
- Currier BH, Currier JH, Franklin KJ, et al. Comparison of Wear and Oxidation in Retrieved Conventional and Highly Cross-Linked UHMWPE Tibial Inserts. J Arthroplasty 2015;30:2349-53. [Crossref] [PubMed]
- Houdek MT, Watts CD, Wyles CC, et al. Metal or Modularity: Why Do Metal-Backed Tibias Have Inferior Outcomes to All-Polyethylene Tibial Components in Patients With Osteoarthritis. J Arthroplasty 2017;32:836-842. [Crossref] [PubMed]
- Berend ME, Ritter MA. The pros and cons of modularity in total knee replacements. J Bone Joint Surg Am 2002;84-A:1480-1; author reply 1. [Crossref] [PubMed]
- O'Keefe TJ, Winter S, Lewallen DG, et al. Clinical and radiographic evaluation of a monoblock tibial component. J Arthroplasty 2010;25:785-92. [Crossref] [PubMed]
- Voigt J, Mosier M. Cemented all-polyethylene and metal-backed polyethylene tibial components used for primary total knee arthroplasty: a systematic review of the literature and meta-analysis of randomized controlled trials involving 1798 primary total knee implants. J Bone Joint Surg Am 2011;93:1790-8. [Crossref] [PubMed]
- Gioe TJ, Stroemer ES, Santos ER. All-polyethylene and metal-backed tibias have similar outcomes at 10 years: a randomized level I [corrected] evidence study. Clin Orthop Relat Res 2007.212-8. [Crossref] [PubMed]
- Bettinson KA, Pinder IM, Moran CG, et al. All-polyethylene compared with metal-backed tibial components in total knee arthroplasty at ten years. A prospective, randomized controlled trial. J Bone Joint Surg Am 2009;91:1587-94. [Crossref] [PubMed]