Comparison of Roche Elecsys and Sysmex HISCL immunoassays for the screening of common blood-borne pathogens
Original Article

Comparison of Roche Elecsys and Sysmex HISCL immunoassays for the screening of common blood-borne pathogens

Wanzhou Xu, Yongqing Tong, Yan Li

Department of Clinical Laboratory, Renmin Hospital of Wuhan University, Wuhan 430060, China

Contributions: (I) Conception and design: Y Li; (II) Administrative support: Y Li; (III) Provision of study materials or patients: W Xu, Y Tong; (IV) Collection and assembly of data: W Xu, Y Tong; (V) Data analysis and interpretation: W Xu; (VI) Manuscript writing: All authors; (VII) Final approval of manuscript: All authors.

Correspondence to: Yan Li. Department of Clinical Laboratory, Renmin Hospital of Wuhan University, 238 Jiefang Road, Wuhan 430060, China. Email: yanlitf1120@163.com.

Background: We conducted a comprehensive comparison in the sensitivity, specificity, dilution sensitivity, and precision between two immunoassay systems, Roche Elecsys Cobas e 601 and Sysmex HISCL 5000, for the detection of hepatitis B virus surface antigen (HBsAg), antibody to hepatitis C virus (anti-HCV), treponemal antibodies (anti-TP), and human immunodeficiency virus (HIV).

Methods: One thousand unselected samples and 100 reserved weak reactive samples were tested by the two systems. Sensitivity and specificity were then calculated for each system. Seroconversion panels were used to assess the sensitivity in early stage detection. Dilution sensitivity was evaluated by dilution tests of several seroconversion panel samples. Evaluation of within-run and intermediate precision was conducted following EP-15A protocol.

Results: The consistency rates of the two systems for the detection of four pathogens were all over 99% among unselected samples. Both Elecsys and HISCL were observed to have high sensitivity and specificity in unselected samples and weak reactive samples. Seroconversion panel tests showed that Elecsys could identify positive results earlier than HISCL in HBsAg (1 out of 4), anti-TP (1 out of 2) and HIV (1 out of 12) panels. The results of the anti-HCV panels were comparable. In dilution tests, Elecsys could detect lower concentrations than HISCL in two anti-HCV samples and one anti-TP sample. The two systems had similar performance in dilution tests of HBsAg and HIV samples. Both Elecsys and HISCL had qualified intra-assay and inter-assay precision.

Conclusions: Both Elecsys and HISCL have good performance in the screening of four common bloodborne pathogens. The two systems are comparable and considered adequate for clinical use.

Keywords: Blood-borne pathogens; serological test; diagnostic performance; Roche Elecsys; Sysmex HISCL


Submitted Jan 29, 2019. Accepted for publication May 17, 2019.

doi: 10.21037/atm.2019.05.83


Introduction

Blood-borne infection is a major concern in transfusion procedures and for health care workers with occupational exposure to patients’ blood (1,2). In recent decades, the human immunodeficiency virus (HIV), hepatitis B virus (HBV), hepatitis C virus (HCV), and syphilis have been the main threats of transfusion-transmitted infections worldwide (3).

In China, blood screening for these pathogens is mandatory for blood donors and essential for most patients under hospitalization. This policy started in the 1990s and effectively decreased the incidence of transfusion-transmitted infection in the next decade (4,5). However, a recent study has reported that the overall positive rates of transfusion-transmitted infection were still 2.11% among 154,038 blood donors in southwestern China (6). This may be due to the window period between onset of infection and completion of seroconversion when serum concentrations of specific antigens or antibodies are insufficient for detection assays to identify a positive result (6,7). Therefore, detection assays of high sensitivity and strong ability in early stage detection are crucial to avoid false-negative results during blood screening.

Currently, enzyme-linked immunosorbent assay (ELISA), electrochemiluminescence immunoassay (ECLIA) and chemiluminescence immunoassay (CLIA) are the most widely used assays in the screening of blood-borne pathogens in China (8). ELISA is a proven method characterized by rapid detection, high sensitivity, and low costs. However, it also has several limitations including inherent instability, lack of automation, and the “Hook effect” (9). ECLIA and CLIA are emerging detection technologies with characteristics of superior diagnostic performance, high precision, good repeatability, and full automation. Xu et al. (10) evaluated the performance of ECLIA and ELISA on HBV detection in 359 serum samples. The results showed that ECLIA had better sensitivity than ELISA in weak reactive samples. Research by Wang et al. (11) indicated that ECLIA could detect positive results earlier than ELISA during seroconversion of HIV infection. Similar results were identified for comparison between CLIA and ELISA (12,13).

Despite these studies, little research has been done to compare ECLIA and CLIA in detecting these blood-borne pathogens, especially for the performance of early-stage detection. The present study aims to compare the comprehensive diagnostic performance of immunoassays from two commercially available detection systems, Roche Elecsys Cobas e 601 (Roche Diagnostics, Penzberg, Germany) (ECLIA) and Sysmex HISCL 5000 (Sysmex Corporation, Kobe, Japan) (CLIA), in the detection of HBV surface antigen (HBsAg), antibody to HCV (anti-HCV), antibody to treponemal antibodies (anti-TP), and HIV (anti-HIV 1/2 and HIV-1 p24Ag).


Methods

Clinical samples

One thousand unselected samples and 100 reserved weak reactive samples were collected from residual blood samples of hospitalized patients in Renmin Hospital of Wuhan University on the condition that they met the following criterion: (I) sample volume >1,000 µL; (II) a sample matrix of serum, heparin plasma, or EDTA plasma; (III) age of subject ≥18. Weak reactive samples were identified with Roche Elecsys system and were defined by the cut off index (COI) for each pathogen marker: COI ranges of 0.8< COI <4.0, 0.9< COI <5.0, and 0.5< COI <5.0 were respectively used for the detection of HBsAg, anti-HCV, and anti-TP. The study protocol was in accordance with the Helsinki Declaration and was approved by the Ethical Committee of Renmin Hospital of Wuhan University.

Seroconversion panels

A total of 29 seroconversion panels (SeraCare Life Sciences) were used to evaluate the sensitivity of the two systems in early stage detection (Table S1). Five panels were used for HBsAg detection, 10 for anti-HCV, 2 for anti-TP, and 12 for HIV (anti-HIV-1 and HIV-1 antigen p24).

Table S1
Table S1 Results of seroconversion panel tests
Full table

Methodologies

Evaluation of sensitivity and specificity

Each clinical sample was separately tested by the two systems for the detection of four pathogen markers. All clinical samples were masked with a unique blind serial number. For inconsistent results, anti-HCV, HIV, and anti-TP samples were confirmed by Mikrogen Immunoblot Tests or recombinant immunobinding assay (RIBA) method; HBsAg samples were tested by a confirmatory test of each system at first, and then Abbott Architect HBsAg assay (Abbott Laboratories, Abbott Park, IL, USA) was used for final confirmation if necessary (Figure 1). All tests for a single clinical sample were completed within 24 hours. The concordance rate, sensitivity, and specificity of each system were then calculated.

Figure 1 Algorithm used to test clinical samples.

Each seroconversion panel was simultaneously tested by Elecsys and HISCL, and the number of days needed to identify a positive result was recorded.

Evaluation of dilution sensitivity

Simulations of gradient concentration sample were completed by diluting seroconversion panels with negative serum into a series of titers (1:5, 1:20, 1:40, 1:80, …, up to 1:163,840). These simulations were sequentially tested by the two systems until the result turned negative.

Evaluation of precision

Within-run and intermediate precision were assessed according to the CLSI EP15-A guideline (14). Concentrations of high level, medium level, and low-level samples for each pathogen marker were tested by Elecsys and HISCL 3 times a day and repeated for 5 days. Low and medium levels were defined according to precision control standards of Roche Elecsys system. High-level samples were selected from serum samples with high COI value (>100). Coefficients of variation (CV%) were calculated.

Detection systems and immunoassays

Roche Elecsys Cobas e 601 (Roche Diagnostics) and Sysmex HISCL 5000 (Sysmex Corporation) are both fully-automated immunoassay analyzers capable for simultaneous detection of antigens and antibodies in serum or plasma, but their mechanisms in detecting antigen-antibody complex are different. For the Cobas e 601, the immune complex attached to the microparticles is magnetically sequestered on the electrode, and the chemiluminescence is electrically induced and measured. Meanwhile, for the HISCL 5000, demagnetization separation technology is used to separate the immune complex and free impurities and chemiluminescence is then measured by filter switching technology.

The immunoassays used for each pathogen are listed in Table 1. Most of them were qualitative assays with the exception of HISCL HBsAg.

Table 1
Table 1 List of immunoassays used for each pathogen marker
Full table

Results

The concordance rates of Elecsys and HISCL were higher than 99% for all pathogen markers in the 1,000 unselected clinical samples (Table 2). A sensitivity of 96.59% for HBsAg and 100% for both anti-HCV and anti-TP was obtained for Elecsys, while the corresponding values were 97.73%, 94.12%, and 100% for HISCL. No positive sample was identified for HIV detection. Both Elecsys and HISCL had an excellent specificity of higher than 99%. Similar results were observed for 100 weak reactive samples (Table 3).

Table 2
Table 2 Detection results of clinical unselected samples by Elecsys and HISCL
Full table
Table 3
Table 3 Detection results of weak reactive samples by Elecsys and HISCL
Full table

Results of seroconversion panel tests are listed in Table S1. Elecsys could identify a positive result earlier than HISCL in three panels: one HBsAg panel (PHM940), one anti-TP panel (PSS901), and one HIV panel (PRB964). However, the results of the 10 anti-HCV panels were controversial: three panels (PHV919, PHV925, and PHV926) favored Elecsys, while three other panels (PHV913, PHV920, and PHV922) favored HISCL.

In the dilution tests of certain seroconversion panel samples, Elecsys could detect titers up to 1:201 and 1:5,120 for two HBsAg panels (PHM937 and PHM939), 1:320 and 1:2,560 for two anti-HCV panels (PHV917 and PHV919), and 1:1,280 for one anti-TP panel (0820-0214-19). The corresponding results for HISCL were 1:40 and 1:20,480 for HBsAg panels, 1:5 and 1:20 for anti-HCV panels, and 1:160 for the anti-TP panel. The two systems could detect an equal dilution ratio (1:40) for the HIV panel PRB976.

Results of precision evaluation showed that both Elecsys and HISCL had a stable performance in all tests in spite of the concentration levels. Within-run and intermediate CV% ranged from 0% to 6.8%, with the results meeting the requirements of the manufacturer.


Discussion

The screening of blood-borne pathogens is crucial to avoid transfusion-transmissible infections and cross infections in hospital, yet the requirements for optimal detection methods are different in two scenarios (15). For screening in blood donations, sensitivity is the most important property to identify every single sample that is potentially infected. In addition, the capability of early-stage detection is also important since infections caused by these pathogens could be asymptomatic and hard to detect in the window period. However, for regular screening in hospitalized patients, minimizing false-positive results should be considered in order to save medical resources and patients’ utility.

We conducted a comprehensive comparison between Roche Elecsys and Sysmex HISCL in their detection performance on blood-borne pathogens, with reference to ECLIA and CLIA respectively. Results of the unselected clinical samples and weak reactive samples indicated that both Elecsys and HISCL had excellent sensitivity and specificity, and the concordance rate of the two systems exceeded 95% in all tests. Additionally, they were fully-automatic analyzers with good within-run and intermediate precision. Thus, our results suggest that both Elecsys and HISCL are capable of screening blood-borne pathogens, regardless of the property concerned.

These results are consistent with previous studies. Tao et al. (16) screened 13,767 serum samples from 13 centers in China and obtained a sensitivity of 100% and a specificity of 99.81% for Elecsys in detecting anti-TP. Similar data on sensitivity and specificity of HISCL was reported by An et al. (17). Elecsys anti-HCV II assay showed a sensitivity of 100% and a specificity of 99.66% in 7,726 routine samples from the Asia-Pacific region (18), while research by Feng et al. (19) revealed that HISCL anti-HCV assay had a sensitivity of 98.97% and specificity of 100% in 1,048 samples from China. Other studies also indicated that both Elecsys and HISCL had superior performance in the detection of HBsAg and HIV (20-22).

Results of seroconversion panel tests demonstrated that Elecsys had a slightly better performance than HISCL in early detection of all pathogen markers except for anti-HCV. Interestingly, Feng et al. (19) compared anti-HCV assays of these two systems and found that Elecsys had a superior seroconversion sensitivity to HISCL (detection rate: 64.62% for Elecsys and 46.15% for HISCL). This advantage may be explained by the difference between ECLIA and CLIA in the detection mechanism. ECLIA yields a specific chemiluminescence reaction initiated by electrochemistry on the surface of the electrode, which is easier to control and more accurate than when initiated by simple mixing of the compound (CLIA) (23). Similar advantages were also observed in dilution tests showing that Elecsys could detect lower titers in several seroconversion panels of HBsAg, anti-HCV, and anti-TP. However, further studies involving samples from different populations are needed to confirm these findings.

Our study had some limitations. First, clinical samples used in our study were retrospectively collected from residual samples of hospitalized patients. Therefore, clinical diagnosis information was not available and could not be used to confirm positive results. Secondly, we did not collect or restrict the genotypes of these pathogens, and so further comparison based on subtypes was not possible. Lastly, no positive HIV samples were collected in our study.

Nevertheless, to the best of our knowledge, this was the first study to compare multiple immunoassays of Elecsys and HISCL in the screening of four common blood-borne pathogens. Our results suggested that both Elecsys and HISCL had superior performance in the detection of HBsAg, anti-HCV, anti-TP, and HIV. The two systems are both adequate for clinical use, but Elecsys probably may have some advantages in early stage detection due to the features of ECLIA.


Acknowledgments

This study was supported by Roche diagnosis (Shanghai) Limited. The funder was also responsible for the supplement of detection reagents and for the support of corresponding detection technology.


Footnote

Conflicts of Interest: The authors have no conflicts of interest to declare.

Ethical Statement: The study protocol was in accordance with the Helsinki Declaration and was approved by the Ethical Committee of Renmin Hospital of Wuhan University.


References

  1. Makroo RN, Hegde V, Chowdhry M, et al. Seroprevalence of infectious markers & their trends in blood donors in a hospital based blood bank in north India. Indian J Med Res 2015;142:317. [Crossref] [PubMed]
  2. Mazi W, Senok AC, Assiri AM, et al. Occupational exposure to blood-borne pathogens in a tertiary hospital: benchmarking using patient days. Asia Pac J Public Health 2015;27:NP1727-32. [Crossref] [PubMed]
  3. Tafesse TB, Gebru AA, Gobalee S, et al. Seroprevalence and diagnosis of HIV, HBV, HCV and syphilis infections among blood donors. Hum Antibodies 2017;25:39-55. [Crossref] [PubMed]
  4. Wang J, Liu J, Yao F, et al. Prevalence, incidence, and residual risks for transfusion-transmitted human immunodeficiency virus Types 1 and 2 infection among Chinese blood donors. Transfusion 2013;53:1240-9. [Crossref] [PubMed]
  5. Gao Z, Zhang Y, Shan H, et al. A 30-year systematic review and meta-analysis of hepatitis B virus among blood donors in mainland China: revealing increase of new threats. Transfusion 2017;57:1988-97. [Crossref] [PubMed]
  6. Xu T, Yi ZM, Luo JM, et al. Prevalence and trends of transfusion-transmittable infections among blood donors in Southwest China. J Public Health (Oxf) 2019;41:55-61. [Crossref] [PubMed]
  7. Kim MJ, Park Q, Min HK, et al. Residual risk of transfusion-transmitted infection with human immunodeficiency virus, hepatitis C virus, and hepatitis B virus in Korea from 2000 through 2010. BMC Infect Dis 2012;12:160. [Crossref] [PubMed]
  8. Fei CR, Ye AQ, Zhang J. Evaluation of different methods in determination of low level HBsAg. Zhejiang Da Xue Xue Bao Yi Xue Ban 2011;40:436-39. [PubMed]
  9. Kift RL, Messenger MP, Wind TC, et al. A comparison of the analytical performance of five commercially available assays for neutrophil gelatinase-associated lipocalin using urine. Ann Clin Biochem 2013;50:236-44. [Crossref] [PubMed]
  10. Xu W, Li Y, Wang M, et al. Comparison of two immunoassays for determining hepatitis B virus serum markers. Clin Chem Lab Med 2011;50:153-7. [PubMed]
  11. Wang T, Li D, Yan K, et al. Performance evaluation of a new fourth-generation HIV Ag/Ab combination electrochemiluminescence immunoassay - evaluation of a new HIV assay. Int J STD AIDS 2014;25:267-72. [Crossref] [PubMed]
  12. Li L, Cai B, Tao C, et al. Performance Evaluation of CLIA for Treponema Pallidum Specific Antibodies Detection in Comparison with ELISA. J Clin Lab Anal 2016;30:216-22. [Crossref] [PubMed]
  13. Farooq A, Waheed U, Zaheer HA, et al. Detection of HBsAg mutants in the blood donor population of Pakistan. PLoS One 2017;12:e0188066. [Crossref] [PubMed]
  14. The National Committee for Clinical Laboratory Standards. User Demonstration of Performance for Precision and Accuracy; Approved Guideline. NCCLs EP-15, 2001.
  15. Kim S, Kim JH, Yoon S, et al. Clinical performance evaluation of four automated chemiluminescence immunoassays for hepatitis C virus antibody detection. J Clin Microbiol 2008;46:3919-23. [Crossref] [PubMed]
  16. Tao C, Hao X, Xu W, et al. Evaluation of the Elecsys syphilis immunoassay for routine screening of serum samples in China. Sci Rep 2017;7:9559. [Crossref] [PubMed]
  17. An J, Chen Q, Liu Q, et al. Evaluation of the HISCL Anti-Treponema pallidum Assay as a Screening Test for Syphilis. Clin Vaccine Immunol 2015;22:817-22. [Crossref] [PubMed]
  18. Yoo SJ, Wang LL, Ning HC, et al. Evaluation of the Elecsys((R)) Anti-HCV II assay for routine hepatitis C virus screening of different Asian Pacific populations and detection of early infection. J Clin Virol 2015;64:20-7. [Crossref] [PubMed]
  19. Feng S, Wei B, Liu Q, et al. Evaluation of the Novel HISCL Chemiluminescence Enzyme Immunoassay for Laboratory Screening of Hepatitis C Virus. Clin Vaccine Immunol 2016;23:652-4. [Crossref] [PubMed]
  20. Gencay M, Seffner A, Pabinger S, et al. Detection of in vivo hepatitis B virus surface antigen mutations-A comparison of four routine screening assays. J Viral Hepat 2018;25:1132-8. [Crossref] [PubMed]
  21. Feng S, Wei B, Rao C, et al. Clinical Evaluation of the Newly Developed HISCL-5000 Analyzer on Detection of Hepatitis B Virus Markers in West China Hospital. Clin Lab 2016;62:1053-60. [Crossref] [PubMed]
  22. Liang S, Deng G, Zhou S, et al. A retrospective analysis of the application of the Elecsys® HIV combi PT assay in southern China. J Clin Lab Anal 2018. [Crossref] [PubMed]
  23. Blackburn GF, Shah HP, Kenten JH, et al. Electrochemiluminescence detection for development of immunoassays and DNA probe assays for clinical diagnostics. Clin Chem 1991;37:1534-39. [PubMed]
Cite this article as: Xu W, Tong Y, Li Y. Comparison of Roche Elecsys and Sysmex HISCL immunoassays for the screening of common blood-borne pathogens. Ann Transl Med 2019;7(14):300. doi: 10.21037/atm.2019.05.83