MR-guided lumbar facet radiofrequency denervation for treatment of patients with chronic low back pain in an open 1.0 Tesla MRI system
Introduction
Chronic low back pain is one of the most commonly diagnosed conditions in developed countries with an enormous socioeconomic burden. In many patients, the condition is attributed to the facet joint pain syndrome with or without radiologically proven spondylarthrosis (1). For patients with a history of frustrating analgetic and physical therapy, corticoid/analgesic infiltration of affected facet joints represents an adequate approach for supplementary diagnosis and first-line minimally invasive treatment (2). In patients who respond to infiltration, treatment can be extended to denervation of the facet joints to achieve a sustained therapeutic effect (3,4). Facet joint radiofrequency denervation (FRD) using fluoroscopy or computed tomography (CT) for image guidance is an established method (5,6).
However, there are several technical issues with fluoroscopy- and CT-guided FRD such as the radiation exposure for patient and interventionalist and poor soft tissue contrast. Higher radiation may be caused by difficult access to the nerve course of the medial dorsal ramus due to degeneratively enlarged facet portions, which can make denervation difficult in some cases. The lack of full three-dimensional angulation of the CT image plane to monitor the position of the thermal applicator may limit the extent of medial dorsal ramus denervation and can thus result in insufficient facet denervation.
Successful magnetic resonance imaging (MRI)-guided interventions in the area of the spinal column have been reported for several indications, such as spinal injection and laser disc decompression (7-13). However, MRI has not yet been used for radiofrequency denervation of the facet joints. In this study, we therefore investigated the feasibility and clinical outcome of multiplanar MRI-guided lumbar FRD in an open 1.0 Tesla MRI scanner.
This article has been written in accordance with the MDAR reporting checklist (available at https://dx.doi.org/10.21037/atm-21-633).
Methods
Study design/patient cohort
This prospective, exploratory single-center study was conducted in accordance with the Declaration of Helsinki (as revised in 2013), approved by the institutional ethics commitee of our university hospital (EA1/071/09, EA1/301/12) and written informed consent was obtained from all patients.
Inclusion criteria: only patients with a preinterventional lumbar pain score of at least 6 on a visual analoge scale (VAS), spondylarthrosis detected by MRI (≥ Fujiwara 2–3°), and multiple positive facet joint infiltrations (>75% pain relief) were included (14). At least 2 of the following 4 symptoms of facet joint pain syndrome were present: local pressure and knocking pain over one or more facets, worsening of low back pain with hyperextension and rotation of the lumbar spine, morning stiffness in the lumbar spine or increased pain in the morning with pseudoradicular radiation into the buttocks or thigh. Exclusion criteria were: progressive neurological deficits due to neuroforaminal and spinal canal stenosis, evidence of lumbosacral radiculopathy, spinal and zygapophyseal cysts, spondylolisthesis, pre- or postinterventional surgery of the lumbar spine, BMI >40 kg/m2, skin infection, circulatory disorders, allergic reactions to local anesthesia, and relevant neurological, cardiological or malignant disease. Indications for facet joint denervation were jointly established by orthopedic surgeons and interventional radiologists.
MRI system and MR-compatible interventional equipment
All interventions were performed in an open 1.0-Tesla MRI system with vertical orientation of the main magnetic field (Panorama HFO, Philips, Best, Netherlands). The open configuration allows almost 360° patient access in the magnetic isocenter and thus provides good conditions for MR-image-guided interventions. The equipment used in the open MRI system included a workstation with complete scanner control, an MRI-compatible in-room monitor, and an MRI-compatible wireless PC mouse for starting interventional acquisitions and switching imaging planes. A flexible solenoid superficial coil (Multipurpose L, Philips, Best, Netherlands) with a diameter of 21 cm was used for all procedures.
All facet denervations were performed using a bipolar radiofrequency (RF) system (CelonLab POWER, Celon AG medical instruments, Teltow, Germany) with a maximum output power of 250 W and a frequency of 470 kHz. The MRI-compatible RF applicator (CelonProSurge MR 150-T20, Celon AG medical instruments, Teltow, Germany) has an electrode length of 20 mm, a shaft length of 150 mm, and a diameter of 15 gauge (1.79 mm).
Preclinical analysis
As no presets for use of the MR-compatible RF applicator for facet joint denervation were avaible at the time of the study, we first carried out ablation tests in an ex vivo porcine model to identify the optimal power, energy, and application duration to be selected in relation to the size of the ablation area to be achieved. For this purpose, radiofrequency ablations at 5 W and different target energy inputs of 100–200 J (in 20 J steps) were combined with different application durations, and the ablation areas/shapes achieved in the animal muscle were subjectively evaluated by two experienced interventionalists in consensus. The aim was to determine the optimal energy input regarding area/shape in terms of denervation effectiveness and maximum protection of surrounding tissue such as paravertebral muscles.
Clinical application
Denervation treatment was performed on an outpatient basis. The interventions were uneventful, and all patients left the hospital after circulatory monitoring for one hour after the procedure.
The interventional team included an MRI technician for coil positioning and pre- and postinterventional imaging, a nurse for sterile patient and equipment preparation, an interventional radiologist, and an orthopedic surgeon. Patients were positioned in 90° lateral position, and the lumbar interventional area was covered sterilely in the usual manner (Figure 1). In order to achieve optimal image quality, the radiofrequency coil was positioned as close as possible to the lumbar puncture site in a position orthogonal to the main magnetic field (B0).
The MRI sequence protocol is presented in Table 1. Multislice T1- and T2-weighted (w) fast spin echo (FSE) sequences were used for preprocedural planning and localization of the target anatomy (Figure 2). An interactive proton density (PD)-w FSE sequence was used to determine the exact skin entry point using the finger-point technique and to direct the applicator needle toward the target anatomy in near real-time (acquisition time: 2 s) (Figure 2). A local anesthetic (2 mL prilocaine 1%, Xylonest™ 1%, AstraZeneca, Wedel, Germany) was administered subcutaneously in the area of the predetermined puncture sites and around the facet joints to be treated.
Full table
After a small skin incision, the RF applicator was positioned parallel to the medial branch of the respective dorsal ramus in approx. 15° inferosuperior puncture direction with posterolateral access. Here, the medial dorsal ramus is located in a groove along the medial-posterior surface of the transverse process (15). For upward advancement of the RF applicator, a puncture site approximately one level below the target site was used. First, the needle should be used to establish bone contact with the transition of the lateral surface of the superior articular process and the transverse process. After this step, the probe was brought into the final position—parallel to the nerve.
For the RFA procedure, a 2W test application was carried out for approx. 2–3 s to rule out irritation of the nerve root after applicator positioning. Each facet joint was treated until the defined autostop settings of the RF generator were reached, and a strongly T2-w fat-saturated SPIR sequence was acquired for assessment of postinterventional effects.
Technical feasibility and safety
Image quality of real-time MRI was evaluated subjectively regarding accurate and safe RF applicator guidance by two experienced interventional radiologists in consensus. The procedure was defined as technically successful when the applicator reached the corresponding target position at the level of the respective medial dorsal ramus branch with MR fluoroscopy for guidance, and FRD was accomplished. Complications were recorded according to the Society of Interventional Radiology (SIR) clinical practice guidelines (16).
Clinical outcome
Back pain was evaluated clinically using follow-up questionnaires with a VAS for subjective assessment of low back pain ranging from 0 (0= no pain) to 10 (10= maximum pain intensity). VAS scores were obtained before the intervention and at follow-up after one week and six months.
Volumetric assessment
To complement the results of subjective pain assessment, we measured paravertebral muscle volumes and fat content. Diagnostic MRI examinations using the same protocols were performed for intervention planning and for follow-up at 6-month intervals. All volumes were measured manually by an experienced reader, and results over time were compared.
The volumetric measurements were performed as described before by Hartwig et al. using Amira v.5.2.0 (Visage Imaging, San Diego, CA, USA) (17). The craniocaudal extent of volumetric measurement was defined according to the treated segments following the definition of Hartwig et al. were the middles of intervertebral discs in the sagittal plane were used to define the superior and inferior margins of the region of interest (17). For example, if segment L4/5 was treated, the region of interest was defined from the center of the L3/4 disc to the center of the L5/S1 disc. The measurements were than performed in the axial T2w with 3mm slice thickness.
Statistical analysis
Graphics were created with GraphPad Prism v.5® (GraphPad Software, San Diego, CA, USA). Statistical analysis was performed with IBM SPSS Statistics 26 for Windows 10 (IBM Corp., Armonk, NY, USA). Descriptive evaluation of the data was done using common measures of location (median, mean), dispersion measures [standard deviation (SD), range from minimum (min) to maximum (max), interquartile range (IQR)], and frequency tables. For exploratory data analysis, we used the Wilcoxon signed-rank test (interval scaled data) and Friedman`s two-way analysis of variance by ranks (ordinal scaled data). Statistical significance was assumed for P<0.05, and Bonferroni correction was used.
Results
Patient population
In this study, 17 patients (7 women/10 men; median age: 58 years, range: 37–83 years) with chronic lumbar facet joint pain syndrome underwent MRI-guided facet joint radiofrequency denervation. Due to the dual nerve supply of each individual segment by descending fibers, the adjacent cranial segment was always treated as well. Segments L2/3 (n=1), L3/4 (n=5) and L4/5 (11) were primary targets. In 11 patients, two segments (n=66 ablations) and in four patients, three segments (n=32 ablations) were denervated in the same session, resulting in a total of 106 facet joint ablations. Further patient details are summarized in Table 2.
Full table
Technical feasibility and safety
Optimal ablation settings identified experimentally were: 5W, 100J, and 20s application duration, resulting in an oval ablation area of 4 mm × 7 mm.
All procedures were technically successful. No major complications occurred during or after the interventions. Mild injection site pain was observed in all patients.
Image quality of the interactive PD-w FSE sequence was adequate, allowing identification of the targeted anatomical landmarks and confirmation of correct needle positioning parallel to the medial dorsal ramus in all patients.
Clinical outcome
Subjective pain scores were improved after the intervention. Preinterventional initial scores (median: 8, IQR: 1, range: 6–9, CI: 7.14–8.04) decreased significantly both after one week (median: 4, IQR: 5, range: 0–7, CI: 1.9–4.69, P=0.003) and after 6 months (median: 1, IQR: 6, range: 0–7, CI: 1.06–4.23, P<0.001) (Figure 3).
Volume measurement
Mean multifidus muscle volume before/after the intervention increased significantly in our patient population (366.8±130.8 cm3/435.4±146.7 cm3, P=0.031). No significant differences were observed for the other volumes analyzed (Figure 4).
Discussion
This proof-of-principle study describes the feasibility of real-time MRI-guided bipolar radiofrequency denervation of lumbar facet joints in patients with chronic lumbar pain syndrome in an open 1.0-Tesla MRI system. Facet joint pain is thought to be mediated by the neural supply of the facet joint capsule. The joints have a rich innervation arising from the medial and lateral branches of the dorsal rami with dual innervation from the medial branches arising from the posterior rami (18). Studies have shown that image-guided denervation of the facets is a safe and effective therapeutic option when indicated and can prevent or postpone surgical treatment (2). In addition to alcohol ablation, radiofrequency denervation using X-ray fluoroscopy or CT guidance has become an established treatment option (3,4,19). Significant pain reduction after denervation for chronic lumbar pain syndrome was reported for a 6-month interval (19) and is confirmed by the results of our study.
Previous studies suggest that MRI-assisted facet infiltration is safe and effective (7-9). The lack of ionizing radiation is advantageous for both patients and interventionalists. Furthermore, MRI navigation has additional advantages. For example, the better soft tissue contrast and especially the multiplanar imaging capabilities of MRI make it easier to position the RF applicator compared to fluoroscopy and CT. MR imaging allows 15° inferosuperior angling of the RF applicator with complete visualization of the applicator. This allows a higher contact length to the medical dorsal ramus for optimized denervation.
We achieved technically successful MR-guided RF denervation of the facet joints in all patients without major complications. The real-time PD-w FSE sequence enabled precise multiplanar applicator positioning with good visualization of spinal structures and optimal artifact properties for applicator imaging. This was accomplished by a complex interaction of material and alignment of the applicator to the main magnetic field (B0), the sequence type, and parameters applied for the vertically oriented magnetic field strength of 1.0 Tesla (7). In comparison to CT, MRI depicts spinal anatomy in greater detail and facilitates applicator positioning due to muliplanar navigation with unlimited selection of image planes (20). In our study, interventions were performed with patients in lateral position, which allows the best possible coil positioning and hence optimal signal amplification (7). In addition, an MRI quadrupole imaging coil (butterfly) was developed for spinal interventions in a vertical 1.0-T MRI system and the use of this surface coil may allow spinal interventions to be performed in a prone position comparable to CT (21). Prone positioning ensures access to a larger area for spinal interventions and might be more familiar to interventionalists, especially to those with little to moderate experience. The durations of interventions were not recorded in this study. The interventionalist performing the procedures estimated the time to be comparable to that of fluoroscopy- or CT-guided denervation after an initial learning curve, due to real-time MR fluoroscopy navigation. Furthermore, the development of dedicated sequences can improve the benefit and safety of MRI-guided interventions (22). Real-time thermometry was reported to technically optimize MRI-guided thermoablation (11,12,23-25). Other technical advances may enable new applications in the future (26,27).
The use of a bipolar RF applicator contributed to the safety of the interventions performed in this study (28). Known risks such as current flow across a patient-neutral electrode (monopolar systems) are prevented by the use of a coaxial electrode (29). Current flow in a bipolar system solely occurs in the distal portion of the applicator. The fact that very low power is required for effective denervation further reduces the risk of unwanted thermal damage. Finally, the autostop function of the applicator which we defined in a preclinical experiment prevents damage to adjacent subcutaneous and muscle tissue as well as to structures at risk such as the nerve root.
A disadvantage of MRI-guided interventions is the currently still limited availability of open MRI scanners. However, techniques for performing interventional MRI guidance in closed-bore systems have been reported. In short wide-bore systems, needles can be advanced with the extended arm using real-time imaging. In standard magnets, control and workflow may be improved by remote operation using robotic or manual driving elements (30).
Regarding clinical outcome, we observed significant pain relief (median VAS scores of 8 to 1), which correlated with an increase in muscle volume (multifidus muscle) after 6 months. The underlying mechanism for this observation is that pain can lead to a sparing of the corresponding segments with subsequent hypotrophy of muscles and increasing fat content (31). Consequently, successful pain therapy is expected to reverse hypotrophy (32). This underlines the benefit of MRI-guided FRD with prompt pain relief (median VAS scores of 8 to 4 one week after denervation), allowing the patient to undergo optimized physiotherapy aimed at improving spinal muscle volume and ensuring sustained pain control.
Besides ethanol and radiofrequency denervation, there are other treatment options such as MRI-guided focused ultrasound ablation or cryoneurolysis that have shown promising results in animals and humans (33,34). Studies comparing different procedures would be desirable and should be performed in the future.
Our study has some limitations. First, we used subjective pain scores without further verification. In addition, patients were asked to report the response to treatment by telephone (not blinded), which could have affected outcome results. Second, a follow-up period of 6 months is relatively short, as pain is bound to recur due to nerve regeneration, and a longer follow-up period would be helpful to analyze the duration of the therapeutic effect and the time when patients may need a repeat procedure. Third, as this was a preliminary study, we did not include a control group to compare MRI guidance with fluoroscopy or CT. Fourth a prospective case number planning was not performed because valid data on the expected effect size are available for CT-guided RFA, but it was not possible to predict with certainty whether comparable effects could be achieved with the MRI-guided procedure. However, the rough estimation of the effect size for this procedure, which is possible from the data of this pilot study, can be used to prospectively generate a sufficient number of cases at adequate test power in following studies. Fifth, this study was performed in an open MRI system, which is not the most common type available. Nevertheless, our experience is transferable to state-of-the-art wide- and short-bore tunnel systems, which are most appropriate for performing MR interventions. For this transfer, some requirements should be considered, such as modifications in patient and coil positioning and possible parameter modifications of the interactive MR sequence. Future studies in larger patient populations should investigate the clinical significance of the achieved pain reduction and long-term efficacy.
In conclusion, this proof-of-principle study shows that MRI-guided facet joint radiofrequency denervation (FRD) in an open 1.0-Tesla MRI system is a promising treatment option for patients with chronic lumbar pain syndrome.
Acknowledgments
The authors thank Bettina Herwig for professional English editing.
Funding: None.
Footnote
Reporting Checklist: The authors have completed the MDAR reporting checklist. Available at https://dx.doi.org/10.21037/atm-21-633
Data Sharing Statement: Available at https://dx.doi.org/10.21037/atm-21-633
Conflicts of Interest: All authors have completed the ICMJE uniform disclosure form (available at https://dx.doi.org/10.21037/atm-21-633). The authors have no conflicts of interest to declare.
Ethical Statement: The authors are accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved. The study was conducted in accordance with the Declaration of Helsinki (as revised in 2013). The study was approved by the institutional ethics commitee of Charité – Universitätsmedizin Berlin (EA1/071/09, EA1/301/12), and informed consent was obtained from all study participants.
Open Access Statement: This is an Open Access article distributed in accordance with the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 International License (CC BY-NC-ND 4.0), which permits the non-commercial replication and distribution of the article with the strict proviso that no changes or edits are made and the original work is properly cited (including links to both the formal publication through the relevant DOI and the license). See: https://creativecommons.org/licenses/by-nc-nd/4.0/.
References
- Urits I, Burshtein A, Sharma M, et al. Low Back Pain, a Comprehensive Review: Pathophysiology, Diagnosis, and Treatment. Curr Pain Headache Rep 2019;23:23. [Crossref] [PubMed]
- Sehgal N, Shah RV, McKenzie-Brown AM, et al. Diagnostic utility of facet (zygapophysial) joint injections in chronic spinal pain: a systematic review of evidence. Pain Physician 2005;8:211-24. [Crossref] [PubMed]
- Leggett LE, Soril LJ, Lorenzetti DL, et al. Radiofrequency ablation for chronic low back pain: a systematic review of randomized controlled trials. Pain Res Manag 2014;19:e146-53. [Crossref] [PubMed]
- Joo YC, Park JY, Kim KH. Comparison of alcohol ablation with repeated thermal radiofrequency ablation in medial branch neurotomy for the treatment of recurrent thoracolumbar facet joint pain. J Anesth 2013;27:390-5. [Crossref] [PubMed]
- Juch JNS, Maas ET, Ostelo R, et al. Effect of Radiofrequency Denervation on Pain Intensity Among Patients With Chronic Low Back Pain: The Mint Randomized Clinical Trials. JAMA 2017;318:68-81. [Crossref] [PubMed]
- Lee CH, Chung CK, Kim CH. The efficacy of conventional radiofrequency denervation in patients with chronic low back pain originating from the facet joints: a meta-analysis of randomized controlled trials. Spine J 2017;17:1770-80. [Crossref] [PubMed]
- Streitparth F, Walter T, Wonneberger U, et al. Image-guided spinal injection procedures in open high-field MRI with vertical field orientation: feasibility and technical features. Eur Radiol 2010;20:395-403. [Crossref] [PubMed]
- Streitparth F, De Bucourt M, Hartwig T, et al. Real-time MR-guided lumbosacral periradicular injection therapy using an open 1.0-T MRI system: an outcome study. Invest Radiol 2013;48:471-6. [Crossref] [PubMed]
- Freyhardt P, Hartwig T, De Bucourt M, et al. MR-guided facet joint injection therapy using an open 1.0-T MRI system: an outcome study. Eur Radiol 2013;23:3296-303. [Crossref] [PubMed]
- Streitparth F, Althoff C, Jonczyk M, et al. Tailored interactive sequences for continuous MR-image-guided freehand biopsies of different organs in an open system at 1.0 tesla (T) - Initial experience. Biomed Tech (Berl) 2017;62:557-63. [Crossref] [PubMed]
- Streitparth F, Hartwig T, Walter T, et al. MR guidance and thermometry of percutaneous laser disc decompression in open MRI: an initial clinical investigation. Eur Radiol 2013;23:2739-46. [Crossref] [PubMed]
- Streitparth F, Walter T, Wonneberger U, et al. MR guidance and thermometry of percutaneous laser disc decompression in open MRI: an ex vivo study. Cardiovasc Intervent Radiol 2014;37:777-83. [Crossref] [PubMed]
- Namba H, Kawasaki M, Izumi M, et al. Effects of MRgFUS Treatment on Musculoskeletal Pain: Comparison between Bone Metastasis and Chronic Knee/Lumbar Osteoarthritis. Pain Res Manag 2019;2019:4867904 [Crossref] [PubMed]
- Fujiwara A, Tamai K, Yamato M, et al. The relationship between facet joint osteoarthritis and disc degeneration of the lumbar spine: an MRI study. Eur Spine J 1999;8:396-401. [Crossref] [PubMed]
- Lau P, Mercer S, Govind J, et al. The surgical anatomy of lumbar medial branch neurotomy (facet denervation). Pain Med 2004;5:289-98. [Crossref] [PubMed]
- Sacks D, McClenny TE, Cardella JF, et al. Society of Interventional Radiology clinical practice guidelines. J Vasc Interv Radiol 2003;14:S199-202. [Crossref] [PubMed]
- Hartwig T, Streitparth F, Gross C, et al. Digital 3-dimensional analysis of the paravertebral lumbar muscles after circumferential single-level fusion. J Spinal Disord Tech 2011;24:451-4. [Crossref] [PubMed]
- Bogduk N. Clinical anatomy of the lumbar spine and sacrum. 4th ed. New York: Churchhill Livingstone; 2005.
- McCormick ZL, Marshall B, Walker J, et al. Long-Term Function, Pain and Medication Use Outcomes of Radiofrequency Ablation for Lumbar Facet Syndrome. Int J Anesth Anesth 2015;2:028.
- Stojanovic MP, Sethee J, Mohiuddin M, et al. MRI analysis of the lumbar spine: can it predict response to diagnostic and therapeutic facet procedures? Clin J Pain 2010;26:110-5. [Crossref] [PubMed]
- Rump JC, Streitparth F, Boning G, et al. Evaluation of a MR-quadrupole imaging coil for spinal interventions in a vertical 1.0 T MRI. Magn Reson Med 2012;68:600-5. [Crossref] [PubMed]
- Makowski MR, Jonczyk M, Streitparth F, et al. Interactive near-real-time high-resolution imaging for MR-guided lumbar interventions using ZOOM imaging in an open 1.0 Tesla MRI system--initial experience. Biomed Tech (Berl) 2015;60:533-9. [Crossref] [PubMed]
- Bazrafshan B, Koujan A, Hubner F, et al. A thermometry software tool for monitoring laser-induced interstitial thermotherapy. Biomed Tech (Berl) 2019;64:449-57. [Crossref] [PubMed]
- Wonneberger U, Schnackenburg B, Wlodarczyk W, et al. Evaluation of thermometric monitoring for intradiscal laser ablation in an open 1.0 T MR scanner. Int J Hyperthermia 2010;26:295-304. [Crossref] [PubMed]
- Wonneberger U, Schnackenburg B, Wlodarczyk W, et al. Intradiscal temperature monitoring using double gradient-echo pulse sequences at 1.0T. J Magn Reson Imaging 2010;31:1499-503. [Crossref] [PubMed]
- Hensen B, Kagebein U, Gutberlet M, et al. Wireless video transmission into the MRI magnet room: implementation and evaluation at 1.5T, 3T and 7T. Biomed Tech (Berl) 2019;64:373-82. [Crossref] [PubMed]
- Schmidt M, Krug JW, Rosenheimer MN, et al. Filtering of ECG signals distorted by magnetic field gradients during MRI using non-linear filters and higher-order statistics. Biomed Tech (Berl) 2018;63:395-406. [Crossref] [PubMed]
- Bruners P, Lipka J, Gunther RW, et al. Bipolar radiofrequency ablation: is the shape of the coagulation volume different in comparison to monopolar RF-ablation using variable active tip lengths? Minim Invasive Ther Allied Technol 2008;17:267-74. [Crossref] [PubMed]
- Burton CV. Percutaneous radiofrequency facet denervation. Appl Neurophysiol 1976-1977;39:80-6. [PubMed]
- Busse H, Kahn T, Moche M. Techniques for Interventional MRI Guidance in Closed-Bore Systems. Top Magn Reson Imaging 2018;27:9-18. [Crossref] [PubMed]
- Goubert D, Oosterwijck JV, Meeus M, et al. Structural Changes of Lumbar Muscles in Non-specific Low Back Pain: A Systematic Review. Pain Physician 2016;19:E985-E1000. [PubMed]
- Bierry G, Kremer S, Kellner F, et al. Disorders of paravertebral lumbar muscles: from pathology to cross-sectional imaging. Skeletal Radiol 2008;37:967-77. [Crossref] [PubMed]
- Weeks EM, Platt MW, Gedroyc W. MRI-guided focused ultrasound (MRgFUS) to treat facet joint osteoarthritis low back pain--case series of an innovative new technique. Eur Radiol 2012;22:2822-35. [Crossref] [PubMed]
- Kaye EA, Monette S, Srimathveeravalli G, et al. MRI-guided focused ultrasound ablation of lumbar medial branch nerve: Feasibility and safety study in a swine model. Int J Hyperthermia 2016;32:786-94. [Crossref] [PubMed]